Company Logo

Uniformity measurements

Two different uniformity test equipments are used, one that uses a Xenon lamp and a set of interferential filters to select the wavelength (see Figure 6) and the other that use he QE measurements facility to obtain a monochromatic beam (see figure 7). An integrating sphere is used to obtain an uniform illumination over a large area in both equipments.

uniform1
Figure 6. Apparatus for uniformity test and measurement using a Xenon lamp and a set of interferential filters.

 

uniform2
Figure 7. Apparatus for uniformity test and measurement using the QE measurements facility.